SBOS690A July 2016 – December 2019 OPA2626
PRODUCTION DATA.
The circuit shown in Figure 55 measures the transient response of the OPA2626. Configure V+, V–, RISO, RLOAD, and CLOAD as desired. Monitor the input and output voltages on an oscilloscope or other signal analyzer. Use this circuit to measure large-signal and small-signal transient response, slew rate, overshoot, and capacitive-load stability.