This mode uses internal averaging to
provide better performance at the tradeoff of higher power consumption.
Procedure
- In the bypass mode box of the
GUI, ensure that the settings match those shown in Figure 3-2
- Once these settings are verified,
press the Device Bringup button.
- Wait until after the message
Device Bringup Completed appears in the Log.
- Under the Analog and Clock
tab, toggle Dither Enable to the off position.
- In HSDC Pro, press the
Capture button. You should now see something similar to Figure 3-3.
- Adjust signal generator output
such that the measured fundamental power in HSDC Pro is at the user's desired
level. (shown below is -12 dBFs)