Cuffless blood pressure monitoring on a wearable device is based on the concept of Pulse Wave Velocity (PWV) which refers to the velocity of the pulse wave across the arteries due to cardiac activity. Blood pressure determines arterial stiffness which in turn affects the PWV. Therefore, changes in blood pressure are reflected as changes in the PWV. Today’s wrist-worn devices support both photoplethysmogram (PPG) and electrocardiogram (ECG) signal acquisitions. Using the relative delay between the ECG and PPG waveforms, a parameter called Pulse Arrival Time (PAT) can be determined which has a correlation to PWV.
In a smartwatch, the PPG signal can be acquired using light-emitting diodes (LEDs) and photodiode (PD) both facing the wrist. The ECG signal can be acquired using a pair of electrodes – one in contact with the left wrist, and the other touched by the finger of the right hand. A third right leg drive (RLD) electrode typically in contact with the wrist, sets the potential relative of the body to the watch ground.
Figure 1-1 shows synchronized ECG and PPG waveforms. PEP, PTT and PAT refer to pre-ejection period, pulse transit time, and pulse arrival time, respectively.
Figure 1-2 shows the scheme for blood pressure monitoring on a wearable device.
The AFE4950 is an analog front end from Texas Instruments that enables synchronized acquisition of PPG and ECG signals, and is designed for cuffless blood pressure monitoring on a wearable device.
AFE4950 OverviewTable 1-1 lists the specifications for a cuffless blood pressure monitoring application.
System Specifications | AFE4950 | Comments |
---|---|---|
Multi-sensor PPG support | 8 LEDs, 4 PDs | Combining signals from multiple PDs and LEDs improves accuracy and helps mitigate motion artifacts. |
Number of ECG Electrodes | 3 | Two electrodes for ECG signal acquisition, and a third electrode to drive RLD |
PPG signal acquisition – peak signal-to-noise ratio (SNR) | 110 dB | Enables high-accuracy PPG monitoring even in low perfusion cases |
ECG signal acquisition - Noise | 0.7 μVRMS | In a 0.5-Hz to 150-Hz band |
Figure 1-3 shows the reference schematic for a cuffless blood pressure monitoring system using AFE4950.
Figure 1-4 shows the AFE4950 signal chain. The output of the PPG and ECG front ends are converted by the same analog-to-digital converter (ADC) in different time slots, resulting in a flawlessly-synchronized signal acquisition of PPG and ECG.
Other analog front ends from Texas Instruments that support blood pressure monitoring on wearable devices include the AFE4960, AFE4960P, and AFE4500.
TI PROVIDES TECHNICAL AND RELIABILITY DATA (INCLUDING DATASHEETS), DESIGN RESOURCES (INCLUDING REFERENCE DESIGNS), APPLICATION OR OTHER DESIGN ADVICE, WEB TOOLS, SAFETY INFORMATION, AND OTHER RESOURCES “AS IS” AND WITH ALL FAULTS, AND DISCLAIMS ALL WARRANTIES, EXPRESS AND IMPLIED, INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR NON-INFRINGEMENT OF THIRD PARTY INTELLECTUAL PROPERTY RIGHTS.
These resources are intended for skilled developers designing with TI products. You are solely responsible for (1) selecting the appropriate TI products for your application, (2) designing, validating and testing your application, and (3) ensuring your application meets applicable standards, and any other safety, security, or other requirements. These resources are subject to change without notice. TI grants you permission to use these resources only for development of an application that uses the TI products described in the resource. Other reproduction and display of these resources is prohibited. No license is granted to any other TI intellectual property right or to any third party intellectual property right. TI disclaims responsibility for, and you will fully indemnify TI and its representatives against, any claims, damages, costs, losses, and liabilities arising out of your use of these resources.
TI’s products are provided subject to TI’s Terms of Sale (www.ti.com/legal/termsofsale.html) or other applicable terms available either on ti.com or provided in conjunction with such TI products. TI’s provision of these resources does not expand or otherwise alter TI’s applicable warranties or warranty disclaimers for TI products.
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2022, Texas Instruments Incorporated