SBOA510 March 2021 OPA455 , OPA462
The oscilloscope used was a four-channel Tektronix DPO-4034. The 1M Ω||13-pF input settings were used on all channels to tolerate input voltage spikes up to 250 VRMS. The x10 probe attenuation settings were used when measuring the supply rail and output voltages to keep the measurements within the viewable area. It should be noted that this reduces the SNR of the scope by 20 dB because the signal is attenuated, but the system measurement noise is not.