SBOU157 September   2015 OPT8241

 

  1.   Voxel Viewer
    1.     Trademarks
    2. 1 Introduction
      1. 1.1 TI 3D Depth Sensors
      2. 1.2 Operating Principles
      3. 1.3 Voxel Viewer and Voxel SDK
    3. 2 User Interface
      1. 2.1 Menu Bar
        1. 2.1.1 File
        2. 2.1.2 Settings
          1. 2.1.2.1 Camera Profiles
          2. 2.1.2.2 Edit Profiles
          3. 2.1.2.3 Statistics
        3. 2.1.3 Windows
        4. 2.1.4 Help
      2. 2.2 Source Bar
      3. 2.3 Main Viewport
      4. 2.4 Left Pane
        1. 2.4.1 Side Viewports
      5. 2.5 Right Pane
        1. 2.5.1 Frequently Used Parameters Window
        2. 2.5.2 Parameter List Window
      6. 2.6 Bottom Pane
        1. 2.6.1 Watch List and Statistics Window
        2. 2.6.2 Data Flow Diagram Window
        3. 2.6.3 Logs Window
    4. 3 Basic Operations
      1. 3.1 Connecting Devices
      2. 3.2 Adjusting the Settings
      3. 3.3 Visualizing the Data
        1. 3.3.1 Ambient
        2. 3.3.2 Amplitude
        3. 3.3.3 Depth
        4. 3.3.4 Distance
        5. 3.3.5 Phase
        6. 3.3.6 Point Clouds
        7. 3.3.7 Histogram
      4. 3.4 De-Noising
        1. 3.4.1 Temporal Filters
          1. 3.4.1.1 IIR Filter
          2. 3.4.1.2 Median Filter
        2. 3.4.2 Spatial Filters
          1. 3.4.2.1 Smooth Filter
          2. 3.4.2.2 Bilateral Filter
          3. 3.4.2.3 Median Filters
        3. 3.4.3 Recommended Starting Point
    5. 4 Calibration
      1. 4.1 Lens Calibration
      2. 4.2 Frequency Calibration
      3. 4.3 Crosstalk Calibration
      4. 4.4 Nonlinearity Calibration
      5. 4.5 Temperature Calibration
      6. 4.6 Common Phase Calibration
      7. 4.7 Pixel-Wise Calibration
      8. 4.8 Profiles and Calibration
    6. 5 TFC Programming
    7. 6 Summary
    8. 7 References

Common Phase Calibration

Common phase calibration assumes the phase-to-depth relationship is linearized by other calibration by the time it runs. The conversion of phase to depth can then be summarized by the linear equation:

Equation 2. eq01_sbou157.gif

The common phase calibration correlates measured phase with actual depth at a single point (φ0i, d0i). This is done for all pixels indexed by i. Only the center pixel is correlated, and the average of 200 samples (default) is used.

fig32_sbou157.gifFigure 32. Common Phase Calibration Setup
fig33_sbou157.gifFigure 33. Frame Capture for Common Phase Calibration