Hold the board by the
edges and flush the top and bottom using a combination of fresh
isopropyl alcohol (IPA) and fresh DI water. Focus on the input traces,
guard copper, and feedback components.
Scrub contaminants from top side of the EVM.
Hold the board by the
edges with the top layer facing up. Flood sensitive input and feedback
areas with fresh IPA.
Use a toothbrush to
gently scrub for 30–60 seconds. Focus on the areas surrounding P1, P2,
JP2, JP3, SH1, and the area directly right of J1.
Flush the scrubbed areas
with fresh DI water, tilting the board to allow runoff to flow away from
the input areas.
Scrub contaminants from the bottom layer TIA
input.
Hold the board by the
edges with the bottom layer facing up. Flood sensitive input and
feedback areas with fresh IPA.
Use a toothbrush to
gently scrub for 30–60 seconds. Focus on areas surrounding U2, R4, Rf,
Cf, K2, K1, U1, R1, and all input and guard traces.
Flush the scrubbed areas
with fresh DI water, tilting the board to allow runoff to flow away from
the input areas.