Figure 3-1 shows the diagram of EVM bench test setup. Refer to Section 2.2 and Section 2.3 in this user's guide for detailed information of headers, jumpers and test
points. Refer to Section 2.4 for the setup assembly and test procedure. Refer to Section 2.5 for the important notice on safety concerns when handling the EVM.
The test setup consists the following
instruments in addition to the EVM board.
- Power Supply: The input dc
voltage source capable of at least 0-80-V and 2 A.
- Load: The load must be
electronic constant-resistance (CR) or constant-current (CC) mode load, capable
of 0 Adc to 2 Adc up to 20 V. For a no-load input current measurement,
disconnect the electronic load as the load can draw a small residual current.
- Multimeters:
- Voltmeter 1: Input
voltage at VIN+ to VIN–. Set voltmeter to an input impedance of 100
MΩ.
- Voltmeter 2:
Output voltage at VOUT+ to VOUT-. Set voltmeter to an input impedance of
100 MΩ.
- Ammeter 1: Input
current. Set ammeter to 1-second aperture time.
- Ammeter 2: Output
current. Set ammeter to 1-second aperture time
- Oscilloscope: With the
scope set to 20-MHz bandwidth and AC coupling, measure the output voltage ripple
directly across an output capacitor with a short ground lead normally provided
with the scope probe. Place the oscilloscope probe tip on the positive terminal
of the output capacitor, holding the ground barrel of the probe through the
ground lead to the negative terminal of the capacitor. TI does not recommend
using a long-leaded ground connection because this can induce additional noise
given a large ground loop. To measure other waveforms, adjust the oscilloscope
as needed.