SLAU640B April 2019 – March 2023 ADC12DJ5200SE
#GUID-3386F6D4-F3E6-441B-ADD1-884AEE24B83E/T5692721-16 shows the jumper settings and #GUID-3386F6D4-F3E6-441B-ADD1-884AEE24B83E/T5692721-17 shows the LED functionality.
Label | Description | Function |
---|---|---|
J12 | ADC power down jumper. | Installed: ADC powered down. Uninstalled: ADC powered up(default) |
J13 | Select the source for Reference clock signal | Installed: External Reference clock signal is selected(default). Uninstalled: Onboard reference signal (LMK61E2) is selected. |
J18 | When hardware calibration trigger option is enabled. The ADC's calibration routine is can be enable using external signal | Installed: ADC's calibration routine is triggered. Uninstalled: ADC's calibration routine is not triggered(default). |
J19 | Selects the source for SPI signals | Installed: SPI signals from FMC+ connector are controlling the devices on the EVM. Uninstalled: SPI signal from the onboard FTDI IC is controlling the devices on the EVMs |
Label | Function |
---|---|
D1 | High temp indicator |
D2 | High input power on channel A |
D3 | High input power on channel B |