Refer to Section 2.5 of this user's guide for the important notice on safety concerns when handling the EVM.
The following are the recommended instructions for the EVM test setup assembly.
Input Connections
- Prior to connecting the DC input source,
set the current limit of the input supply to 100 mA maximum. Verify the input source is
initially set to 0 V and connected to the VIN+ and VIN– connection points, as shown in
Figure 3-1. An additional input bulk capacitor is recommended to provide damping if long input
lines are used.
- Connect Voltmeter 1 at VIN+ and VIN–
connection points to measure the input voltage.
- Connect Ammeter 1 to measure the input
current and set to at least 1-second aperture time.
Output Connections
- Connect electronic load to VOUT+ and
VOUT- connection points as shown in Figure 3-1. Set the load to constant-resistance mode or constant-current mode at 0 A before
applying input voltage.
- Connect Voltmeter 2 at VOUT+ and VOUT-
connection points to measure the output voltage.
- Connect Ammeter 2 measure the output
current.
Test Procedure
- Set up the EVM as described above.
- Set load to constant resistance or
constant current mode and to sink 10 mA.
- Increase input source from 0 V to 20 V;
use voltmeter 1 to measure the input voltage.
- Increase the current limit of the input
supply to 1000 mA.
- Use voltmeter 2 to measure the output
voltage, vary the load from 10 mA to 1000 mA DC; VOUT must remain within the load
regulation specification.
- Set the load current to 500 mA (50% rated
load) and vary the input source voltage from 20 V to 60 V; VOUT must remain within the
line regulation specification.
- Use the oscilloscope to probe different
signals of interest. Be careful of the probe ground reference.
- Decrease load to 10 mA. Decrease input
source voltage to 0 V.
- Shut down both the input source and load,
and then the voltmeters and ammeters.