SBAS661C February   2015  – May 2021 ADS1262 , ADS1263

PRODUCTION DATA  

  1. Features
  2. Applications
  3. Description
  4. Revision History
  5. Device Comparison
  6. Pin Configuration and Functions
  7. Specifications
    1. 7.1 Absolute Maximum Ratings
    2. 7.2 ESD Ratings
    3. 7.3 Recommended Operating Conditions
    4. 7.4 Thermal Information
    5. 7.5 Electrical Characteristics
    6. 7.6 Timing Requirements: Serial Interface
    7. 7.7 Switching Characteristics: Serial Interface
    8. 7.8 Timing Diagrams
    9. 7.9 Typical Characteristics
  8. Parameter Measurement Information
    1. 8.1 Offset Temperature Drift Measurement
    2. 8.2 Gain Temperature Drift Measurement
    3. 8.3 Common-Mode Rejection Ratio Measurement
    4. 8.4 Power-Supply Rejection Ratio Measurement
    5. 8.5 Crosstalk Measurement (ADS1263)
    6. 8.6 Reference-Voltage Temperature-Drift Measurement
    7. 8.7 Reference-Voltage Thermal-Hysteresis Measurement
    8. 8.8 Noise Performance
  9. Detailed Description
    1. 9.1 Overview
    2. 9.2 Functional Block Diagram
    3. 9.3 Feature Description
      1. 9.3.1  Multifunction Analog Inputs
      2. 9.3.2  Analog Input Description
        1. 9.3.2.1 ESD Diode
        2. 9.3.2.2 Input Multiplexer
      3. 9.3.3  Sensor Bias
      4. 9.3.4  Temperature Sensor
      5. 9.3.5  Power-Supply Monitor
      6. 9.3.6  PGA
      7. 9.3.7  PGA Voltage Overrange Monitors
        1. 9.3.7.1 PGA Differential Output Monitor
        2. 9.3.7.2 PGA Absolute Output-Voltage Monitor
      8. 9.3.8  ADC Reference Voltage
        1. 9.3.8.1 Internal Reference
        2. 9.3.8.2 External Reference
        3. 9.3.8.3 Power-Supply Reference
        4. 9.3.8.4 Low-Reference Monitor
      9. 9.3.9  ADC1 Modulator
      10. 9.3.10 Digital Filter
        1. 9.3.10.1 Sinc Filter Mode
          1. 9.3.10.1.1 Sinc Filter Frequency Response
        2. 9.3.10.2 FIR Filter
        3. 9.3.10.3 50-Hz and 60-Hz Line Cycle Rejection
      11. 9.3.11 Sensor-Excitation Current Sources (IDAC1 and IDAC2)
      12. 9.3.12 Level-Shift Voltage
      13. 9.3.13 General-Purpose Input/Output (GPIO)
      14. 9.3.14 Test DAC (TDAC)
      15. 9.3.15 ADC2 (ADS1263)
        1. 9.3.15.1 ADC2 Inputs
        2. 9.3.15.2 ADC2 PGA
        3. 9.3.15.3 ADC2 Reference
        4. 9.3.15.4 ADC2 Modulator
        5. 9.3.15.5 ADC2 Digital Filter
    4. 9.4 Device Functional Modes
      1. 9.4.1  Conversion Control
        1. 9.4.1.1 Continuous Conversion Mode
        2. 9.4.1.2 Pulse Conversion Mode
        3. 9.4.1.3 ADC2 Conversion Control (ADS1263)
      2. 9.4.2  Conversion Latency
      3. 9.4.3  Programmable Time Delay
      4. 9.4.4  Serial Interface
        1. 9.4.4.1 Chip Select (CS)
        2. 9.4.4.2 Serial Clock (SCLK)
        3. 9.4.4.3 Data Input (DIN)
        4. 9.4.4.4 Data Output/Data Ready (DOUT/DRDY)
        5. 9.4.4.5 Serial Interface Autoreset
      5. 9.4.5  Data Ready Pin (DRDY)
      6. 9.4.6  Conversion Data Software Polling
      7. 9.4.7  Read Conversion Data
        1. 9.4.7.1 Read Data Direct (ADC1 Only)
        2. 9.4.7.2 Read Data by Command
        3. 9.4.7.3 Data-Byte Sequence
          1. 9.4.7.3.1 Status Byte
          2. 9.4.7.3.2 Data Byte Format
          3. 9.4.7.3.3 Checksum Byte (CRC/CHK)
            1. 9.4.7.3.3.1 Checksum Mode (CRC[1:0] = 01h)
          4. 9.4.7.3.4 CRC Mode (CRC[1:0] = 10h)
      8. 9.4.8  ADC Clock Modes
        1. 9.4.8.1 Internal Oscillator
        2. 9.4.8.2 External Clock
        3. 9.4.8.3 Crystal Oscillator
      9. 9.4.9  Calibration
        1. 9.4.9.1 Offset and Full-Scale Calibration
          1. 9.4.9.1.1 Offset Calibration Registers
          2. 9.4.9.1.2 Full-Scale Calibration Registers
        2. 9.4.9.2 ADC1 Offset Self-Calibration (SFOCAL1)
        3. 9.4.9.3 ADC1 Offset System Calibration (SYOCAL1)
        4. 9.4.9.4 ADC2 Offset Self-Calibration ADC2 (SFOCAL2)
        5. 9.4.9.5 ADC2 Offset System Calibration ADC2 (SYOCAL2)
        6. 9.4.9.6 ADC1 Full-Scale System Calibration (SYGCAL1)
        7. 9.4.9.7 ADC2 Full-Scale System Calibration ADC2 (SYGCAL2)
        8. 9.4.9.8 Calibration Command Procedure
        9. 9.4.9.9 User Calibration Procedure
      10. 9.4.10 Reset
        1. 9.4.10.1 Power-On Reset (POR)
        2. 9.4.10.2 RESET/PWDN Pin
        3. 9.4.10.3 Reset by Command
      11. 9.4.11 Power-Down Mode
      12. 9.4.12 Chop Mode
    5. 9.5 Programming
      1. 9.5.1 NOP Command
      2. 9.5.2 RESET Command
      3. 9.5.3 START1, STOP1, START2, STOP2 Commands
      4. 9.5.4 RDATA1, RDATA2 Commands
      5. 9.5.5 SYOCAL1, SYGCAL1, SFOCAL1, SYOCAL2, SYGCAL2, SFOCAL2 Commands
      6. 9.5.6 RREG Command
      7. 9.5.7 WREG Command
    6. 9.6 Register Maps
      1. 9.6.1  Device Identification Register (address = 00h) [reset = x]
      2. 9.6.2  Power Register (address = 01h) [reset = 11h]
      3. 9.6.3  Interface Register (address = 02h) [reset = 05h]
      4. 9.6.4  Mode0 Register (address = 03h) [reset = 00h]
      5. 9.6.5  Mode1 Register (address = 04h) [reset = 80h]
      6. 9.6.6  Mode2 Register (address = 05h) [reset = 04h]
      7. 9.6.7  Input Multiplexer Register (address = 06h) [reset = 01h]
      8. 9.6.8  Offset Calibration Registers (address = 07h, 08h, 09h) [reset = 00h, 00h, 00h]
      9. 9.6.9  Full-Scale Calibration Registers (address = 0Ah, 0Bh, 0Ch) [reset = 40h, 00h, 00h]
      10. 9.6.10 IDACMUX Register (address = 0Dh) [reset = BBh]
      11. 9.6.11 IDACMAG Register (address = 0Eh) [reset = 00h]
      12. 9.6.12 REFMUX Register (address = 0Fh) [reset = 00h]
      13. 9.6.13 TDACP Control Register (address = 10h) [reset = 00h]
      14. 9.6.14 TDACN Control Register (address = 11h) [reset = 00h]
      15. 9.6.15 GPIO Connection Register (address = 12h) [reset = 00h]
      16. 9.6.16 GPIO Direction Register (address = 13h) [reset = 00h]
      17. 9.6.17 GPIO Data Register (address = 14h) [reset = 00h]
      18. 9.6.18 ADC2 Configuration Register (address = 15h) [reset = 00h]
      19. 9.6.19 ADC2 Input Multiplexer Register (address = 16h) [reset = 01h]
      20. 9.6.20 ADC2 Offset Calibration Registers (address = 17h, 18h) [reset = 00h, 00h]
      21. 9.6.21 ADC2 Full-Scale Calibration Registers (address = 19h, 1Ah) [reset = 00h, 40h]
  10. 10Application and Implementation
    1. 10.1 Application Information
      1. 10.1.1 Isolated (or Floated) Inputs
      2. 10.1.2 Single-Ended Measurements
      3. 10.1.3 Differential Measurements
      4. 10.1.4 Input Range
      5. 10.1.5 Input Filtering
        1. 10.1.5.1 Aliasing
      6. 10.1.6 Input Overload
      7. 10.1.7 Unused Inputs and Outputs
      8. 10.1.8 Voltage Reference
      9. 10.1.9 Serial Interface Connections
    2. 10.2 Typical Application
      1. 10.2.1 3-Wire RTD Measurement with Lead-Wire Compensation
        1. 10.2.1.1 Design Requirements
        2. 10.2.1.2 Detailed Design Procedure
        3. 10.2.1.3 Application Curve
    3. 10.3 What To Do and What Not To Do
    4. 10.4 Initialization Setup
  11. 11Power Supply Recommendations
    1. 11.1 Power-Supply Decoupling
    2. 11.2 Analog Power-Supply Clamp
    3. 11.3 Power-Supply Sequencing
  12. 12Layout
    1. 12.1 Layout Guidelines
    2. 12.2 Layout Example
  13. 13Device and Documentation Support
    1. 13.1 Receiving Notification of Documentation Updates
    2. 13.2 Support Resources
    3. 13.3 Trademarks
    4. 13.4 Electrostatic Discharge Caution
    5. 13.5 Glossary

Package Options

Mechanical Data (Package|Pins)
Thermal pad, mechanical data (Package|Pins)
Orderable Information

Serial Interface Autoreset

The CS input resets the serial interface when taken high. However, applications that tie the CS pin low do not have the ability to reset the serial interface by using this pin. If a false SCLK occurs (for example, caused by a noise pulse or clocking glitch), the serial interface may inadvertently advance one or more bit positions, resulting in loss of synchronization to the external microcontroller. If loss of synchronization occurs, the interface does not respond correctly until the interface is reset.

For applications that tie CS low, the ADC provides a feature that automatically resets the serial interface in the event of a glitch. As shown in Figure 9-41, after the first SCLK low-to-high transition is detected by the ADC (either caused by a glitch or a normal SCLK input), if the ADC does not detect seven additional SCLK transitions within 65536 fCLK cycles (approximately 8.9 ms), the serial interface resets. After reset, the interface is ready for the next transaction four fCLK cycles later.

If the seven SCLK transitions are detected within the 65536 fCLK cycles, the serial interface is not reset, and the SCLK detection cycle restarts at the next SCLK transition.

If the serial interface loses synchronization to an external controller, reset the serial interface by holding SCLK low for 65536 fCLK cycles.

The serial interface autoreset function is enabled by the setting TIMEOUT = 1 (bit 3 of the INTERFACE register). The default mode is off.

GUID-4A3B4448-CA05-478B-A644-C1B45812F696-low.gifFigure 9-41 Serial Interface Autoreset
Table 9-16 Autoreset Timing Requirement
PARAMETER TEST CONDITIONS MIN UNIT
td(SCRS) SCLK↑ transition to interface reset : delay time TIMEOUT bit =1 65536 tCLK(1)
td(RSSC) Serial interface reset to first SCLK↑: delay time TIMEOUT bit =1 4 tCLK
tCLK = 1 / fCLK.