AFE4960P

AKTIV

Zweikanaliges EKG- und PPG-Analog-Frontend mit integrierter Atmungs- und Tempoerkennung

Produktdetails

Applications 1-lead ECG, 3-lead ECG, 5-lead ECG, HRV, Optical-HRM, Optical-HRM for wearables, Patient monitoring, Respiration, Sleep study monitor, SpO2, Vo2 max Number of input channels 6 Resolution (Bits) 22 Number of LEDs 4 Interface type Photodiode, Electrodes Operating temperature range (°C) -40 to 85 Rating Catalog
Applications 1-lead ECG, 3-lead ECG, 5-lead ECG, HRV, Optical-HRM, Optical-HRM for wearables, Patient monitoring, Respiration, Sleep study monitor, SpO2, Vo2 max Number of input channels 6 Resolution (Bits) 22 Number of LEDs 4 Interface type Photodiode, Electrodes Operating temperature range (°C) -40 to 85 Rating Catalog
DSBGA (YBG) 36 5.76 mm² 2.4 x 2.4
  • Supports ECG and respiration impedance measurement; configurable as 2-channel ECG or 1-channel ECG + 1-channel respiration
  • Can be used in IEC 60601-2-47:2012/(R)2016 and IEC 60601-2-27:2011(R)2016 compliant systems
  • 222 µA/channel while operating as 2-channel ECG
  • Integrated Pace pulse detection on 1 channel
  • Supports a 3-lead ECG, extendable to 5 leads or higher by operating two or more AFEs in parallel
  • PPG signal chain for SpO2, Heart rate and PTT
  • ECG signal chain:
    • Single channel ECG acquisition up to 2.048 kHz
    • 2 channel ECG acquisition up to 1.36 kHz/channel
    • RLD output to set body bias through third electrode
    • Programmable INA gain from approximately 2 to 12
    • >1 GΩ input impedance, CMRR > 100 dB
    • Input noise (0.5-150 Hz): 13 µVpp at INA gain of 3; 5 µVpp at INA gain of 12
    • Integrated 370 Hz anti-aliasing low pass filter
    • Mode for continuous lead on/off detection
    • Mode for lead impedance measurement
  • Bio-Z signal chain:
    • Measurement of Bio-impedance over excitation frequency from 30 kHz to 100 kHz
    • Sine wave or square wave excitation
    • Respiration impedance measurement: 45 mΩ-pp noise on a baseline impedance of 2 kΩ
  • Dual channel ECG channel:
    • Bio-Z receiver configurable as a 2nd ECG channel
  • PPG Signal Chain
    • Supports acquisition of up to 24 PPG signals
    • Flexible allocation of 4 LED, 4 PD in each phase
    • 8-Bit LED Current control up to 250 mA
    • Low noise current DAC at input to cancel DC
    • Trans-impedance Gain: 3.7 kΩ to 1 MΩ
    • Filter to reduce sensor’s optical noise bandwidth
  • External clock and internal oscillator modes
  • FIFO with 128-sample Depth, 24-bit word
  • SPITM, I2C interfaces: Selectable by pin
  • 2.6-mm × 2.6-mm DSBGA, 0.4-mm Pitch
  • Supplies: Rx:1.7-1.9 V, IO:1.7-1.9 V , Tx: 3.0-5.5 V
  • Supports ECG and respiration impedance measurement; configurable as 2-channel ECG or 1-channel ECG + 1-channel respiration
  • Can be used in IEC 60601-2-47:2012/(R)2016 and IEC 60601-2-27:2011(R)2016 compliant systems
  • 222 µA/channel while operating as 2-channel ECG
  • Integrated Pace pulse detection on 1 channel
  • Supports a 3-lead ECG, extendable to 5 leads or higher by operating two or more AFEs in parallel
  • PPG signal chain for SpO2, Heart rate and PTT
  • ECG signal chain:
    • Single channel ECG acquisition up to 2.048 kHz
    • 2 channel ECG acquisition up to 1.36 kHz/channel
    • RLD output to set body bias through third electrode
    • Programmable INA gain from approximately 2 to 12
    • >1 GΩ input impedance, CMRR > 100 dB
    • Input noise (0.5-150 Hz): 13 µVpp at INA gain of 3; 5 µVpp at INA gain of 12
    • Integrated 370 Hz anti-aliasing low pass filter
    • Mode for continuous lead on/off detection
    • Mode for lead impedance measurement
  • Bio-Z signal chain:
    • Measurement of Bio-impedance over excitation frequency from 30 kHz to 100 kHz
    • Sine wave or square wave excitation
    • Respiration impedance measurement: 45 mΩ-pp noise on a baseline impedance of 2 kΩ
  • Dual channel ECG channel:
    • Bio-Z receiver configurable as a 2nd ECG channel
  • PPG Signal Chain
    • Supports acquisition of up to 24 PPG signals
    • Flexible allocation of 4 LED, 4 PD in each phase
    • 8-Bit LED Current control up to 250 mA
    • Low noise current DAC at input to cancel DC
    • Trans-impedance Gain: 3.7 kΩ to 1 MΩ
    • Filter to reduce sensor’s optical noise bandwidth
  • External clock and internal oscillator modes
  • FIFO with 128-sample Depth, 24-bit word
  • SPITM, I2C interfaces: Selectable by pin
  • 2.6-mm × 2.6-mm DSBGA, 0.4-mm Pitch
  • Supplies: Rx:1.7-1.9 V, IO:1.7-1.9 V , Tx: 3.0-5.5 V

The AFE4960P supports synchronized signal acquisition of ECG, PPG and bio-impedance signals. A PPG signal chain comprises of a current driver for LEDs and a receiver that senses the signal from PDs. The AFE can be configured as either a 2-channel ECG receiver or as a 1-channel ECG receiver and a Respiration impedance channel. The AFE signal chains can interface in a flexible manner to up to 4 electrodes. The AFE has AC and DC Lead biasing and lead on/off detection. Pacemaker pulse detection is supported on one channel. All the signal chain outputs are converted by a single ADC in well-defined time slots and stored as 24-bit words in a 128-sample FIFO which can be read out using a SPI or an I2C interface. The AFE4960P can be used to realize a 3-lead ECG system along with SpO2.

The AFE4960P supports synchronized signal acquisition of ECG, PPG and bio-impedance signals. A PPG signal chain comprises of a current driver for LEDs and a receiver that senses the signal from PDs. The AFE can be configured as either a 2-channel ECG receiver or as a 1-channel ECG receiver and a Respiration impedance channel. The AFE signal chains can interface in a flexible manner to up to 4 electrodes. The AFE has AC and DC Lead biasing and lead on/off detection. Pacemaker pulse detection is supported on one channel. All the signal chain outputs are converted by a single ADC in well-defined time slots and stored as 24-bit words in a 128-sample FIFO which can be read out using a SPI or an I2C interface. The AFE4960P can be used to realize a 3-lead ECG system along with SpO2.

Herunterladen Video mit Transkript ansehen Video

Weitere Informationen anfordern

Das vollständige Datenblatt und weitere Designressourcen sind unter NDA verfügbar. Jetzt anfordern

Ähnliche Produkte, die für Sie interessant sein könnten

Selbe Funktionalität wie der verglichene Baustein bei gleicher Anschlussbelegung
AFE4960 AKTIV Zweikanaliges EKG-Analog-Frontend mit integrierter Atmungs- und Tempoerkennung Similar design without integrated photoplethysmography (PPG) analog front end

Technische Dokumentation

star =Von TI ausgewählte Top-Empfehlungen für dieses Produkt
Keine Ergebnisse gefunden. Bitte geben Sie einen anderen Begriff ein und versuchen Sie es erneut.
Alle anzeigen 9
Typ Titel Datum
* Data sheet AFE4960 Two-Channel ECG, Respiration and Pace Pulse Detection Analog Front End (AFE) for Clinical Wearables datasheet PDF | HTML 13 Mai 2022
Product overview ECG and Cuffless Blood Pressure Monitoring With Integrated AFE PDF | HTML 30 Okt 2023
Application brief Wearable EEG (Electroencephalography) (Rev. B) PDF | HTML 30 Okt 2023
Application brief Remote Patient Monitoring PDF | HTML 30 Nov 2022
Application brief ECG Signal Acquisition on Wearables PDF | HTML 15 Nov 2022
Application brief Measuring Oxygen Saturation (SpO2) on Wearables PDF | HTML 14 Nov 2022
Application brief Optical Heart Rate Monitoring (OHRM) on Wearables PDF | HTML 14 Nov 2022
Application brief Cuffless Blood Pressure Monitoring on Wearables PDF | HTML 09 Nov 2022
Certificate AFE4960PEVM EU RoHS Declaration of Conformity (DoC) 29 Jul 2022

Design und Entwicklung

Weitere Bedingungen oder erforderliche Ressourcen enthält gegebenenfalls die Detailseite, die Sie durch Klicken auf einen der unten stehenden Titel erreichen.

Evaluierungsplatine

AFE4960PEVM — AFE4960P-Evaluierungsmodul für Zweikanal-EKG- und PPG-AFE mit Atmungs- und Tempoerkennung

Das AFE4960P-Evaluierungsmodul (EVM) ist eine Plattform zur Auswertung des AFE4960P-AFE (Biosensing Analog Front End), das der Unterstützung synchronisierter Signalerfassungsfunktionen von Elektrokardiogramm (EKG) und Photoplethysmographie (PPG) dient.

AFE4960P kann entweder als zweikanaliger (...)

Simulationstool

PSPICE-FOR-TI — PSpice® für TI Design-und Simulationstool

PSpice® für TI ist eine Design- und Simulationsumgebung, welche Sie dabei unterstützt, die Funktionalität analoger Schaltungen zu evaluieren. Diese voll ausgestattete Design- und Simulationssuite verwendet eine analoge Analyse-Engine von Cadence®. PSpice für TI ist kostenlos erhältlich und (...)
Gehäuse Pins CAD-Symbole, Footprints und 3D-Modelle
DSBGA (YBG) 36 Ultra Librarian

Bestellen & Qualität

Beinhaltete Information:
  • RoHS
  • REACH
  • Bausteinkennzeichnung
  • Blei-Finish/Ball-Material
  • MSL-Rating / Spitzenrückfluss
  • MTBF-/FIT-Schätzungen
  • Materialinhalt
  • Qualifikationszusammenfassung
  • Kontinuierliches Zuverlässigkeitsmonitoring
Beinhaltete Information:
  • Werksstandort
  • Montagestandort

Empfohlene Produkte können Parameter, Evaluierungsmodule oder Referenzdesigns zu diesem TI-Produkt beinhalten.

Support und Schulungen

TI E2E™-Foren mit technischem Support von TI-Ingenieuren

Inhalte werden ohne Gewähr von TI und der Community bereitgestellt. Sie stellen keine Spezifikationen von TI dar. Siehe Nutzungsbedingungen.

Bei Fragen zu den Themen Qualität, Gehäuse oder Bestellung von TI-Produkten siehe TI-Support. ​​​​​​​​​​​​​​

Videos