AFE4900

現行

適用於穿戴式光學、電子生物感測的超低功耗整合式 AFE

產品詳細資料

Applications 1-lead ECG, Optical-HRM Number of input channels 3 Resolution (Bits) 22 Number of LEDs 4 Interface type I2C, SPI Operating temperature range (°C) -20 to 70 Rating Catalog
Applications 1-lead ECG, Optical-HRM Number of input channels 3 Resolution (Bits) 22 Number of LEDs 4 Interface type I2C, SPI Operating temperature range (°C) -20 to 70 Rating Catalog
DSBGA (YZ) 30 4.8 mm² 2.4 x 2
  • Synchronized PPG, ECG signal acquisition at data rates up to 1 kHz
  • ECG signal chain:
    • Standalone ECG acquisition up to 4 kHz
    • Input bias: 1-lead ECG with RLD bias
    • Programmable INA gain: 2.15 to 12.92
    • Input noise (1 Hz to 150 Hz): 2.5 µVrms at 1 kHz, 1.25 µVrms at 4-kHz data rate
    • AC, DC lead-off detect: 12.5-nA to 100-nA
    • IEC 60601 Test report available on request
  • PPG receiver:
    • Supports three time-multiplexed PD inputs
    • 24-Bit representation of current from PD
    • DC offset subtraction DAC (Up to ±126-µA) at TIA input for each LED, ambient
    • Digital ambient subtraction at ADC output
    • Noise filtering with programmable bandwidth
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Dynamic range up to 100 dB
    • Receiver operates in PPG-only mode at approximately 1-µA/Hz sampling rate
    • Power-down mode: approximately 0 µA
  • PPG transmitter:
    • Four LEDs in common anode configuration
    • 8-Bit LED current up to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of three LEDs for SpO2, or multiwavelength HRM
    • Average current of 30 µA adequate for a
      typical heart-rate monitoring scenario:
      • 20-mA Setting, 60-µs pulse duration,
        25-Hz sampling rate
  • Clocking using an external or internal clock
  • FIFO with 128-sample depth for ECG and PPG
  • I2C, SPI interfaces: selectable by pin
  • 2.6-mm × 2.1-mm, 0.4-mm Pitch DSBGA package
  • Supplies:
    • Rx: 1.8 V to 1.9 V (LDO bypass),
      2.0 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.25 V
    • IO: 1.7 V to Rx_SUP
  • Synchronized PPG, ECG signal acquisition at data rates up to 1 kHz
  • ECG signal chain:
    • Standalone ECG acquisition up to 4 kHz
    • Input bias: 1-lead ECG with RLD bias
    • Programmable INA gain: 2.15 to 12.92
    • Input noise (1 Hz to 150 Hz): 2.5 µVrms at 1 kHz, 1.25 µVrms at 4-kHz data rate
    • AC, DC lead-off detect: 12.5-nA to 100-nA
    • IEC 60601 Test report available on request
  • PPG receiver:
    • Supports three time-multiplexed PD inputs
    • 24-Bit representation of current from PD
    • DC offset subtraction DAC (Up to ±126-µA) at TIA input for each LED, ambient
    • Digital ambient subtraction at ADC output
    • Noise filtering with programmable bandwidth
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Dynamic range up to 100 dB
    • Receiver operates in PPG-only mode at approximately 1-µA/Hz sampling rate
    • Power-down mode: approximately 0 µA
  • PPG transmitter:
    • Four LEDs in common anode configuration
    • 8-Bit LED current up to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of three LEDs for SpO2, or multiwavelength HRM
    • Average current of 30 µA adequate for a
      typical heart-rate monitoring scenario:
      • 20-mA Setting, 60-µs pulse duration,
        25-Hz sampling rate
  • Clocking using an external or internal clock
  • FIFO with 128-sample depth for ECG and PPG
  • I2C, SPI interfaces: selectable by pin
  • 2.6-mm × 2.1-mm, 0.4-mm Pitch DSBGA package
  • Supplies:
    • Rx: 1.8 V to 1.9 V (LDO bypass),
      2.0 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.25 V
    • IO: 1.7 V to Rx_SUP

The AFE4900 device is an analog front-end (AFE) for synchronized electrocardiogram (ECG), photoplethysmogram (PPG) signal acquisition. The device can also be used for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The PPG signal chain supports up to four switching light-emitting diodes (LEDs) and up to three photodiodes (PDs). The LEDs can be switched on using a fully integrated LED driver. The current from the photodiode is converted into voltage by the transimpedance amplifier (TIA) and digitized using an analog-to-digital converter (ADC). The ECG signal chain has an instrumentation amplifier (INA) with a programmable gain that interfaces to the same ADC. A right-leg drive (RLD) amplifier set can be used to the bias for the ECG input pins. AC and dc lead-off detect schemes are supported. The ADC codes from the PPG and ECG phases can be stored in a 128-sample first in, first out (FIFO) block and read out using either an I2C or a serial programming interface (SPI) interface.

The AFE4900 device is an analog front-end (AFE) for synchronized electrocardiogram (ECG), photoplethysmogram (PPG) signal acquisition. The device can also be used for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The PPG signal chain supports up to four switching light-emitting diodes (LEDs) and up to three photodiodes (PDs). The LEDs can be switched on using a fully integrated LED driver. The current from the photodiode is converted into voltage by the transimpedance amplifier (TIA) and digitized using an analog-to-digital converter (ADC). The ECG signal chain has an instrumentation amplifier (INA) with a programmable gain that interfaces to the same ADC. A right-leg drive (RLD) amplifier set can be used to the bias for the ECG input pins. AC and dc lead-off detect schemes are supported. The ADC codes from the PPG and ECG phases can be stored in a 128-sample first in, first out (FIFO) block and read out using either an I2C or a serial programming interface (SPI) interface.

下載 觀看有字幕稿的影片 影片
Request more information

Full data sheet and other design resources are available under NDA. Request now

您可能會感興趣的類似產品

open-in-new 比較替代產品
功能相同,但引腳輸出與所比較的產品不同
AFE4950 現行 適用於穿戴式光學心率監控和 SpO2 的超小型整合式 AFE ECG+PPG AFE with integrated high-pass, low-pass filters optimized for dry electrodes.

技術文件

star =TI 所選的此產品重要文件
找不到結果。請清除您的搜尋條件,然後再試一次。
檢視所有 3
類型 標題 日期
* Data sheet AFE4900 Ultra-low Power, Integrated AFE for Wearable Optical, Electrical Bio-sensing with FIFO datasheet (Rev. B) PDF | HTML 2019年 7月 30日
Technical article Designing compact, efficient and high performing multiparameter patient monitors PDF | HTML 2019年 6月 25日
Technical article No strings attached: Creating next-generation wireless patient monitors PDF | HTML 2018年 6月 29日

設計與開發

如需其他條款或必要資源,請按一下下方的任何標題以檢視詳細頁面 (如有)。

開發板

AFE4403EVM — 適用於心率監控器和脈搏血氧儀的 AFE4403 整合式 AFE 評估模組

The AFE4403EVM is intended for evaluating the AFE4403 device. The fully integrated AFE is ideally suited for Pulse Oximeter and Heart rate monitoring applications and consists of low noise receive channel, the LED transmit section and diagnostics for sensor and LED fault detection. The AFE4403 has (...)

使用指南: PDF
TI.com 無法提供
開發板

AFE4490SPO2EVM — 適用於脈搏血氧儀的 AFE4490 整合式類比前端 (AFE) 評估模組

The AFE44x0SPO2EVM is intended for evaluating AFE4400 and AFE4490 devices. The family of devices are fully integrated AFE ideally suited for Pulse Oximeter applications and consist of low noise receive channel, the LED transmit section and diagnostics for sensor and LED fault detection. The (...)

使用指南: PDF
TI.com 無法提供
開發板

AFE4900EVM — 適用於穿戴式光學、電生物感測的 AFE4900 超低功耗整合式 AFE 評估模組

The AFE4900 is an analog front end (AFE) for optical biosensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The device supports four switching light-emitting diodes (LEDs) and three photodiodes. The current from the photodiode is converted (...)

模擬工具

PSPICE-FOR-TI — PSpice® for TI 設計與模擬工具

PSpice® for TI is a design and simulation environment that helps evaluate functionality of analog circuits. This full-featured, design and simulation suite uses an analog analysis engine from Cadence®. Available at no cost, PSpice for TI includes one of the largest model libraries in the (...)
參考設計

TIDA-01580 — 適用於醫療和消費者穿戴式裝置的無線 ECG、SpO2、PTT 和心率監測器參考設計

This reference design is a simple, wearable, multi-parameter, patient monitor that uses a single-chip, biosensing front end, AFE4900, for synchronized electrocardiography (ECG) and photoplethysmography (PPG) measurements. The measured data is transferred to a remote location using CC2640R2F (...)
Design guide: PDF
電路圖: PDF
封裝 針腳 CAD 符號、佔位空間與 3D 模型
DSBGA (YZ) 30 Ultra Librarian

訂購與品質

內含資訊:
  • RoHS
  • REACH
  • 產品標記
  • 鉛塗層/球物料
  • MSL 等級/回焊峰值
  • MTBF/FIT 估算值
  • 材料內容
  • 認證摘要
  • 進行中持續性的可靠性監測
內含資訊:
  • 晶圓廠位置
  • 組裝地點

建議產品可能具有與此 TI 產品相關的參數、評估模組或參考設計。

支援與培訓

內含 TI 工程師技術支援的 TI E2E™ 論壇

內容係由 TI 和社群貢獻者依「現狀」提供,且不構成 TI 規範。檢視使用條款

若有關於品質、封裝或訂購 TI 產品的問題,請參閱 TI 支援。​​​​​​​​​​​​​​

影片